Visualization of Flow Patterns From Stents and Stent-Grafts in an In Vitro Flow-Model
- 1 October 1998
- journal article
- research article
- Published by Wolters Kluwer Health in Investigative Radiology
- Vol. 33 (10) , 762-770
- https://doi.org/10.1097/00004424-199810000-00008
Abstract
The authors determine flow characteristics and pressure gradients of different stents and stent grafts in an in vitro flow-model. Five vascular stents (Memotherm, Cragg, two Palmaz P308, Strecker, and Wall) and one stent graft (Cragg EndoPro System 1), equal in length (60 mm) and diameter (10 mm), were deployed in a closed flow-model. The inner diameter of the tube measured 9 mm. Flow at 1.5 L/min, 6 L/min, and 10 L/min was simulated. Flow patterns were visualized by anionic particles illuminated with two Helium-Neon lasers. Laminary flow characteristics and pre-/poststent pressure gradients were determined in either expanded stent, 25% stenosis, or 50% stenosis. Stent implantation induced a decrease of laminary flow compared with an unstented tube with and without concentric 25% stenosis (P P The investigated stents and stent grafts showed different severity of flow disturbances and pressure gradients at different graded stenoses. Inadequate stent deployment bears the risk of creating less laminary flow and pathologic pressure gradients. Because flow disturbances and pressure gradients may influence neointimal hyperplasia, stent design and completeness of stent unfolding are important regarding the appearance of postinterventional restenoses.Keywords
This publication has 19 references indexed in Scilit:
- Mittelfristige Ergebnisse nach Cragg-Stent-Implantation in die BeckenarterienRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 1997
- Occlusive and aneurysmal peripheral arterial disease: assessment of a stent-graft system.Radiology, 1996
- Aortic and iliac stenoses: follow-up results of stent placement after insufficient balloon angioplasty in 118 cases.Radiology, 1996
- Chordal geometry determines the shape and extent of systolic anterior mitral motion: In vitro studiesJournal of the American College of Cardiology, 1989
- In vitro flow dynamics of four prosthetic aortic valves: A comparative analysisJournal of Biomechanics, 1989
- The effects of a glycerin-based blood analog on the testing of bioprosthetic heart valvesJournal of Biomechanics, 1989
- In vitro comparison of aortic heart valve prosthesesThe Journal of Thoracic and Cardiovascular Surgery, 1988
- Flow in tubes and arteries - a comparisonBiorheology, 1986
- Fresh Human Aortic and Artificial Heart Valves Studied in Vitro Using Ultramicroscope Anemometry. Further ResultsThe Thoracic and Cardiovascular Surgeon, 1982
- Transluminally-placed Coilspring Endarterial Tube GraftsInvestigative Radiology, 1969