Thermal effects in tissues from combined simultaneous coaxial CO2 and Nd:YAG laser beams

Abstract
This study examines the depth of thermal coagulation and charring in swine liver, kidney cortex, tongue (inferior surface), skeletal muscle, inflated lung, and skin resulting from in vivo incision with simultaneous coaxial CO2 and Nd:YAG (1.064 μM) laser beams. At values of 20 w and 40 w, respectively, and at values of 30 w and 60 w, respectively, of combined CO2 and Nd:YAG laser radiation, coagulation depths determined histologically in liver were significantly greater (PP2 and Nd:YAG laser power were comparable to those obtained by other workers in liver and other relatively vascular tissues using a contact Nd:YAG laser tip. Charring depths obtained at power settings of 30 w (CO2) and 60 w (Nd:YAG) were greater in liver (P2 to Nd:YAG laser power in the range 2–3 in the more vascular tissues, liver and kidney cortex, whereas hemostasis was achieved also in the lesser vascular tissues at higher values. These results strongly suggest the usefulness of combined simultaneous CO2 and Nd:YAG laser beams in surgery of the more vascular organs and tissues.