Scatter factors in proton therapy with a broad beam
- 21 March 2006
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (7) , 1919-1928
- https://doi.org/10.1088/0031-9155/51/7/020
Abstract
In the absence of a predictor of beam output in proton therapy using a broad beam, the beam output is obtained for individual treatments by calibrating the beam monitors. The calibration is carried out under conditions similar to the treatment conditions but with a phantom instead of the patient. However, the dose in the phantom a priori differs from that in the patient. In order to deliver the accurate dose, a correction factor has been introduced to correct the difference. This correction factor is referred to as a scatter factor in an analogy with photon therapy, and is defined as the ratio of the dose at the prescription point in the patient to the dose at the calibration point in the phantom. Under the calibration conditions at Hyogo Ion Beam Medical Center (HIBMC), the range compensator and the collimator, which are usually required in proton therapy with a broad beam, are not used. Therefore the scatter factor includes the effects of the devices as well as the difference between the dose in the patient and that in the phantom. We have developed an estimator using a dose calculation based on the pencil beam algorithm and implemented it in a treatment planning system (TPS) for clinical use. This estimator estimates the scatter factor by calculating the ratio of the doses under the same conditions in the TPS. In order to evaluate the performance of the estimator, demonstrations were carried out for cases with measurable outcomes using a gantry nozzle at HIBMC. We observed 2-3% differences between the measurements and the estimations. These differences were considered to result from the limitations of the dose calculation algorithm in modelling the beam and the patient.Keywords
This publication has 13 references indexed in Scilit:
- Ridge filter design for proton therapy at Hyogo Ion Beam Medical CenterPhysics in Medicine & Biology, 2003
- A CT calibration method based on the polybinary tissue model for radiotherapy treatment planningPhysics in Medicine & Biology, 2003
- Implementation of pencil kernel and depth penetration algorithms for treatment planning of proton beamsPhysics in Medicine & Biology, 1999
- Current Status of Patient Positioning at HIMAC.RADIOISOTOPES, 1998
- Monte Carlo and analytical calculation of proton pencil beams for computerized treatment plan optimizationPhysics in Medicine & Biology, 1997
- A pencil beam algorithm for proton dose calculationsPhysics in Medicine & Biology, 1996
- The calibration of CT Hounsfield units for radiotherapy treatment planningPhysics in Medicine & Biology, 1996
- An empirical method to build up a model of proton dose distribution for a radiotherapy treatment-planning packagePhysics in Medicine & Biology, 1993
- Differential‐pencil‐beam dose calculations for charged particlesMedical Physics, 1992
- Electron beam dose calculationsPhysics in Medicine & Biology, 1981