Application of nonresonance excitation to ion trap tandem mass spectrometry and selected ejection chemical ionization
- 1 July 1996
- journal article
- Published by American Chemical Society (ACS) in Journal of the American Society for Mass Spectrometry
- Vol. 7 (7) , 668-676
- https://doi.org/10.1016/1044-0305(96)00011-6
Abstract
Nonresonance excitation is a universal ion excitation and ejection method in which increased ion kinetic energy is achieved by the combination of an axial dc dipole and the rf trapping fields. The method does not require the applied excitation frequency to match with the secular frequency of the precursor ions to effect collision-induced dissociation (CID) for tandem mass spectrometry applications. Therefore, it is free of the effects of secular frequency changes caused by space-charge and simplifies the optimization of tandem mass spectrometry parameters when combined with gas chromatography-tandem mass spectrometry (GC-MS/MS). Computer simulations show that in contrast to the resonance excitation process, the nonresonance excitation process is able to accelerate thermal ions to kinetic energies in excess of 40 eV in a few microseconds. Based on simulations, we expect that the rapid deposition of energy by this method may allow the study, in ion traps, of high energy decomposition channels of precursor ions with multiple decomposition channels. Furthermore, the method is able to simultaneously excite multiple precursor ions, for example, excite both analyte and its coeluting isotopically labeled internal standard for GC-MS/MS analysis. A GC-MS/MS analysis of 100 pg of n-butylbenzene is demonstrated with a signal-to-noise ratio of 3624, which is over an order of magnitude higher than the signal-to-noise ratio of 345 obtained by full scan gas chromatography-mass spectrometry. In addition, the nonresonance excitation method can be used as a low pass mass filter in the chemical ionization (CI) mode to eject undesired fragment ions that result from direct electron ionization. This new CI method, selected ejection chemical ionization, can produce a CI spectrum without contamination of sample fragment ions from electron ionization.Keywords
This publication has 25 references indexed in Scilit:
- Analysis of neuropeptides by perfusion liquid chromatography/electrospray ion-trap mass spectrometryRapid Communications in Mass Spectrometry, 1994
- Evaluation of storage of ions over m/z 1000 in an ion trap mass spectrometerInternational Journal of Mass Spectrometry and Ion Processes, 1993
- Enhanced mass resolution in a quadrupole ion trapRapid Communications in Mass Spectrometry, 1993
- Improved accuracy of mass measurement with a quadrupole ion‐trap mass spectrometerRapid Communications in Mass Spectrometry, 1992
- Operation of a quadrupole ion trap mass spectrometer to achieve high mass/charge ratiosInternational Journal of Mass Spectrometry and Ion Processes, 1991
- High resolution on a quadrupole ion trap mass spectrometerJournal of the American Society for Mass Spectrometry, 1991
- Electrospray ionization combined with ion trap mass spectrometryAnalytical Chemistry, 1990
- Extending the mass range of the quadrupole ion trap using axial modulationRapid Communications in Mass Spectrometry, 1989
- Injection of ions into a quadrupole ion trap mass spectrometerInternational Journal of Mass Spectrometry and Ion Processes, 1989
- Tandem mass spectrometry of prostaglandins: A comparison of an ion trap and a reversed geometry sector instrumentRapid Communications in Mass Spectrometry, 1988