Tracheal size following tracheostomy with cuffed tracheostomy tubes: an experimental study.
Open Access
- 1 June 1975
- Vol. 30 (3) , 271-277
- https://doi.org/10.1136/thx.30.3.271
Abstract
In view of the severe damage caused by unyielding, low residual volume cuffs, various modifications to the cuff of an intratracheal tube have been introduced. The merits of two low-pressure cuffs were assessed in an experimental study in dogs; both cuffs produced little visible damage to the tracheal wall in dogs intubated continuously over a two-week period. A modified technique of producing tantalum tracheobronchograms without distrubing the mucous blanket or traumatizing the tracheal wall is described. These tantalum radiological studies demonstrated a progressive temporary increase in size of the trachea at cuff level over the period of intubation with these cuffs. The implications of such a progressive weakness occurring in the tracheal muscle are discussed.Keywords
This publication has 21 references indexed in Scilit:
- Pathogenesis of tracheal stenosis following tracheostomy with a cuffed tube: An experimental study in dogsThorax, 1972
- Squeegee of fluid in the compressed canine tracheaRespiration Physiology, 1972
- A New Low-pressure Cuff for Endotracheal TubesAnesthesiology, 1971
- Determination of Persistence of Tantalum Dust In Vivo Following Bronchography, Using Reactor-activated Tantalum and Total Body CountingInvestigative Radiology, 1970
- The Evolution of Tracheal Injury Due to Ventilatory Assistance Through Cuffed TubesAnnals of Surgery, 1969
- Powdered Tantalum as a Medium for Bronchography in Canine and Human LungsInvestigative Radiology, 1968
- A NEW TRACHEOSTOMY TUBEActa Anaesthesiologica Scandinavica, 1967
- Aetiology and repair of tracheal stenosis following tracheostomy and intermittent positive pressure respiration.Thorax, 1967
- Tracheal lesions after tracheostomy.Thorax, 1965
- AUTOMATIC INTERMITTENT INFLATION OF TRACHEOSTOMY-TUBE CUFFThe Lancet, 1964