Electrochemical evaluation of glucose oxidase immobilized by different methods
- 1 March 1975
- journal article
- research article
- Published by Wiley in Biotechnology & Bioengineering
- Vol. 17 (3) , 413-422
- https://doi.org/10.1002/bit.260170309
Abstract
Glucose oxidase electrodes were constructed on a platinum screen using polyacrylamide gel, glutaraldehyde crosslinking, and glutaraldehyde crosslinking with +0.04 volts dc on the platinum screen as the methods of enzyme immobilization. The electrodes were evaluated in an electrochemical cell for the oxidation of glucose at the enzyme electrode and the reduction of oxygen at a platinum auxiliary electrode, using constant current voltametry or under external load operation. The method of immobilization affected the extrapolated opencircuit potential as well as the half‐cell potential and the steady current under external load operation. The charged glutaraldehyde electrode gave the best current performance; however, the small output (microamps) indicated that major problems in electron transfer from an enzyme catalyst to an external circuit must be resolved before such electrodes can be used in practical application.Keywords
This publication has 3 references indexed in Scilit:
- Electrochemical measurements with glucose oxidase immobilized in polyacrylamide gel: Constant current voltametryBiotechnology & Bioengineering, 1971
- Effect of the Microenvironment on the Mode of Action of Immobilized EnzymesPublished by Wiley ,1971
- INTERMOLECULAR CROSS LINKING OF A PROTEIN IN THE CRYSTALLINE STATE: CARBOXYPEPTIDASE-AProceedings of the National Academy of Sciences, 1964