Optimal CT settings for bone evaluations
- 1 May 1985
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 30 (5) , 401-409
- https://doi.org/10.1088/0031-9155/30/5/003
Abstract
In CT densitometry the precision, sensitivity and accuracy of the procedure are limited by photon noise, the degree of tissue differentiation and artefacts caused by the finite size of the X-ray beam, hardening of the X-rays and scattered radiation. These artefacts are highly dependent on the CT settings, such as X-ray spectrum and effective energy, intensity of photon flux and collimation of the beam. To monitor the progression of a disease or the efficacy of a therapy, longitudinal studies are necessary; thus the radiation dose per examination should be kept at a minimum. For bone evaluation of the peripheral skeleton it is shown that there is a narrow optimal energy range of 30-40 keV where a dose below 150 mu Sv is sufficient to detect a change of 1% in trabecular bone density at the 0.1 confidence level. Concomitantly accuracy might be better than 1% if adequate beam collimation is applied.Keywords
This publication has 10 references indexed in Scilit:
- DEVELOPMENT OF QUANTITATIVE CT METHODS FOR BONE AND SOFT TISSUE EVALUATIONSJournal of Computer Assisted Tomography, 1983
- Kleincomputertomograph für quantitative Knochenuntersuchungen in den Extremitäten des Menschen - Special Purpose CT-System for Quantitative Bone Evaluations in the Appendicular SkeletonBiomedizinische Technik/Biomedical Engineering, 1983
- The exponential edge-gradient effect in X-ray computed tomographyPhysics in Medicine & Biology, 1981
- A Framework for Spectral Artifact Corrections in X-Ray CTIEEE Transactions on Biomedical Engineering, 1981
- Optimum energy for performing CT iodinated constrast studiesThe British Journal of Radiology, 1980
- Nonlinear partial volume artifacts in x‐ray computed tomographyMedical Physics, 1980
- Blurring in Tomograms Made with X-Ray Beams of Finite WidthJournal of Computer Assisted Tomography, 1979
- Statistical limitations in x‐ray reconstructive tomographyMedical Physics, 1976
- Scatter Considerations in Fan Beam Computerized Tomographic SystemsIEEE Transactions on Nuclear Science, 1976
- The Fourier reconstruction of a head sectionIEEE Transactions on Nuclear Science, 1974