In vitro evaluation of relative perforating potential of central venous catheters: Comparison of materials, selected models, number of lumens, and angles of incidence to simulated membrane
- 1 January 1991
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 7 (1) , 1-6
- https://doi.org/10.1007/bf01617891
Abstract
Perforation of the vena cava or atrium is a serious complication of monitoring with a central venous catheter. We designed an in vitro model with a pulsating simulated membrane to evaluate a number of variables that could affect relative perforating potential of different types of central venous catheters. To determine the perforating potential of central venous catheters, we studied the effects of (1) the angle of incidence (n=6) between catheter and simulated membrane; (2) catheter material (polyurethane and polyethylene); (3) make (manufacturer and model) (n=6), with 3 catheters of each make tested; (4) design (n=3 each: silicone rubber, open-ended, blunt-ended, and polyurethane pigtail); and (5) number of lumens (single, double, or triple). Each trial was repeated five times with each catheter that was tested. Perforation was significantly more likely when the angle of incidence between catheter and pulsating simulated membrane was greater than 40° than when it was 40° (PPPP<0.001). These data offer additional objective information to consider when choosing and positioning central venous catheters.Keywords
This publication has 25 references indexed in Scilit:
- Cardiac Tamponade from Central Venous Catheters Report of a Case and Review of the English LiteratureAngiology, 1984
- Cardiac tamponade associated with a multilumen central venous catheterCritical Care Medicine, 1984
- Material Thrombogenicity in Central Venous Catheterization III. A Comparison Between Soft Polyvinylchloride and Soft Polyurethane Elastomer, Long, Antebrachial CathetersActa Anaesthesiologica Scandinavica, 1984
- Material Thrombogenicity in Central Venous Catheterization II. A Comparison between Plain Silicone Elastomer, and Plain Polyethylene, Long, Antebrachial CathetersActa Anaesthesiologica Scandinavica, 1983
- Stiffness of Central Venous CathetersActa Anaesthesiologica Scandinavica, 1983
- Fatal cardiac tamponade. Occurrence with other complications after left internal jugular vein catheterizationJAMA, 1982
- How Much do Arm Movements Displace Cubital Central Venous Catheters?Acta Anaesthesiologica Scandinavica, 1982
- Cardiac Tamponade From Central Venous CathetersArchives of Surgery, 1982
- Cardiac Tamponade from Central Venous CathetersAnesthesiology, 1979
- Mechanism of perforation of the heart with production of hydropericardium by a venous catheter and its preventionThe American Journal of Surgery, 1970