On the Determination of an Optimum Microwave Diathermy Frequency for a Direct Contact Applicator
- 1 April 1966
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. BME-13 (2) , 76-87
- https://doi.org/10.1109/TBME.1966.4502410
Abstract
An approximate theoretical analysis is presented for determining the relative heating in a two layer fat-muscle medium due to a dielectric loaded dipole-corner reflector applicator in direct contact with the finite fat layer. The results indicate that convenient applicator sizes, equivalent to those now used at 2450 Mc/s, operating at a frequency in the neighborhood of 750 Mc/s provide the maximum ratio of relative heat per unit volume in the muscle as compared to that in the fat. Experimental data taken with the aid of an electrical model of the fat-muscle layers shows reasonable agreement with the results of the approximate analysis.Keywords
This publication has 4 references indexed in Scilit:
- The complete electromagnetic field of a half-wave dipole in a dissipative mediumIEEE Transactions on Antennas and Propagation, 1963
- The Mechanism of Absorption of Ultrahigh Frequency Electromagnetic Energy in Tissues, as Related to the Problem of Tolerance DosageIRE Transactions on Medical Electronics, 1956
- THE BIOPHYSICAL BASIS OF PHYSICAL MEDICINEJAMA, 1956
- Heating of fat-muscle layers by electromagnetic and ultrasonic diathermyTransactions of the American Institute of Electrical Engineers, Part I: Communication and Electronics, 1953