Early Changes in Defibrillation Threshold Following Implantation of a Nonthoracotomy System in Dogs
- 1 November 1994
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 17 (11) , 1771-1777
- https://doi.org/10.1111/j.1540-8159.1994.tb03744.x
Abstract
Nonthoracotomy systems are rapidly becoming the preferred surgical method for implantation of cardioverter defibrillators. Testing is performed at the time of implantation to insure an adequate margin of safety for defibrillation. However, this safety margin may change with lead maturation. This study evaluated changes in defibrillation threshold following implantation of a nonthoracotomy system. Ten dogs underwent implantation of a nonthoracotomy system consisting of a single catheter with a distal coil electrode in the right ventricular apex and a proximal coil electrode in the superior vena cava forming a common anode with a subcutaneous patch over the left thorax. Defibrillation threshold testing, using a biphasic waveform, was performed on each animal under general anesthesia at implantation (day 1) and subsequently on postoperative days 3, 7, 10, 17, 24, 31, 38, and 45. E50, the energy associated with a 50% likelihood of successful defibrillation, was determined at each setting. The mean E50 was 12.2 +/- 1.1 J at the time of implantation, increasing 36% to 16.8 +/- 2.0 J by day 38 (P < 0.01). Individual increases in E50 of 10-12 J were observed in four animals. Energy requirements for defibrillation with a nonthoracotomy system increase during the early postoperative period, with the highest defibrillation threshold observed at 38 days. This increase may be applicable to humans and should be considered when selecting an adequate energy safety margin for defibrillation at time of implantation.Keywords
This publication has 18 references indexed in Scilit:
- Implantable Cardioverter Defibrillator Implanted by Nonthoracotomy Approach: Initial Clinical Experience with the Redesigned Transvenous Lead SystemPacing and Clinical Electrophysiology, 1991
- The Effect of Multiple Shocks on Canine Cardiac DefibrillationPacing and Clinical Electrophysiology, 1990
- An effective and adaptable transvenous defibrillation system using the coronary sinus in humansJournal of the American College of Cardiology, 1990
- Defibrillation threshold testing: Necessary but evil?American Heart Journal, 1989
- Success of chronic defibrillation and the role of antiarrhythmic drugs with the automatic implantable cardioverter/defibrillatorThe American Journal of Cardiology, 1987
- The relationship between successful defibrillation and delivered energy in open-chest dogs: Reappraisal of the “defibrillation threshold” conceptAmerican Heart Journal, 1987
- Catheter electrode defibrillation in dogs: Threshold dependence on implant time and catheter stabilityAmerican Heart Journal, 1985
- Lead systems for internal ventricular fibrillation.Circulation, 1981
- Ventricular defibrillation in the dog using implanted and partially implanted electrode systemsThe American Journal of Cardiology, 1974