Electrical properties of commercial concentric EMG electrodes
- 1 January 1985
- journal article
- research article
- Published by Wiley in Muscle & Nerve
- Vol. 8 (1) , 1-8
- https://doi.org/10.1002/mus.880080102
Abstract
Five electrical characteristics—impedance, broadband noise generation, line interference sensitivity, signal distortion, and common‐mode conversion—were measured in five electromyographic (EMG) concentric needle electrodes (CNEs) from each of six commercial manufacturers. Untreated CNEs showed considerable variation in impedance and broadband noise characteristics, both within and among manufacturers. Electrolytic treatment reduced impedances by a factor between 1.5 and 4.0 and lessene within manufacture variability Average post‐treatment impedances at 100Hz ranged from 31 to 436 kOhms, reflecting in part the range of core surface areas. Treatment also reduced the broadband noise to the level of the instrumentation noise for all but the highest impedance CNEs. Distortion and common‐mode conversion were negligible for the lowest impedance CNEs. Line interference from a nearby power cord was completely suppressed only by those CNEs with fully shielded cables, and then only when the electromyographer also was grounded; there was no measurable benefit when the shield was driven, as opposed to grounded. We conclude that (1) there are consistent differences in the properties of CNEs from different manufacturers, reflecting differences in materials, design, and construction; (2) electrolytic treatment temporarily improves the performance of all CNEs, and tends to lessen the differences among them, most likely through a surface electrochemical effect; and (3) both active and passive cable shields are effective in reducing extraneous line interference.Keywords
This publication has 7 references indexed in Scilit:
- A Unified Approach to Represent Metal Electrode PolarizationIEEE Transactions on Biomedical Engineering, 1983
- Linear and nonlinear properties of platinum electrode polarisation II: time domain analysisMedical & Biological Engineering & Computing, 1983
- Linear and nonlinear properties of platinum electrode polarisation. Part 1: frequency dependence at very low frequenciesMedical & Biological Engineering & Computing, 1982
- Linear AC Electrode Polarization Impedance at Smooth Noble Metal InterfacesIEEE Transactions on Biomedical Engineering, 1977
- Report of the committee on EMG instrumentation: Technical factors in recording electrical activity of muscle and nerve in manElectroencephalography and Clinical Neurophysiology, 1970
- Action Potential Parameters in Normal Human Muscle and their Dependence on Physical Variables.Acta Physiologica Scandinavica, 1954
- The discharge of impulses in motor nerve fibresThe Journal of Physiology, 1929