The desflurane Tec 6 vaporizer
Open Access
- 1 April 1994
- journal article
- research article
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 72 (4) , 470-473
- https://doi.org/10.1093/bja/72.4.470
Abstract
A late development model of the Tec 6 vaporizer for the administration of desflurane vapour has been evaluated, It is heated electrically and has both electronic monitors of vaporizer function and alarms. The new filling system is a significant improvement over previous Tec filling systems. The vaporizer requires a warm-up period before it may be used, but when activated it provides an output that is approximately linear between 1 and 18% vapour concentrations, at flow rates between 200 ml min−1 and 10 litre min−1. (Br. J. Anaesth. 1994; 72: 470–473)Keywords
This publication has 0 references indexed in Scilit: