Detection of hyperoxaemia in neonates: data from three new pulse oximeters
Open Access
- 1 November 2002
- journal article
- research article
- Published by BMJ in Archives of Disease in Childhood: Fetal & Neonatal
- Vol. 87 (3) , 217F-219
- https://doi.org/10.1136/fn.87.3.f217
Abstract
Aim: To determine the sensitivity and specificity of three newly developed pulse oximeters in the detection of hyperoxaemia, defined as an arterial partial pressure of oxygen (Pao2) of > 80 mm Hg. Methods: Spo2 readings from three oximeters (Agilent Viridia (AgV), Masimo SET (MaS), Nellcor Oxismart (NeO)) were documented in 56 infants (median gestational age at birth 35.5 weeks, range 24–41) whenever an arterial blood gas was taken for clinical purposes. Blood samples were analysed within one minute in a Radiometer ABL 505 blood gas analyser and OSM3 co-oximeter. Results: Between 280 and 291 blood gases were analysed for each instrument; 105–112 showed a Pao2 > 80 mm Hg. At an upper alarm limit of 95%, the three instruments detected hyperoxaemia with 93–95% sensitivity. Specificity at this alarm level ranged from 26 to 45%. The mean (SD) difference between arterial oxygen saturation and Spo2 (bias) was −0.25 (2.5)% for AgV, −0.06 (2.5)% for MaS, and −0.91 (2.6)% for NeO (p < 0.01, NeO v AgV and MaS). Conclusion: These instruments detected hyperoxaemia with sufficient sensitivity at an upper alarm limit of 95%, but showed differences in their specificity, which was probably related to differences in measurement bias.Keywords
This publication has 13 references indexed in Scilit:
- Oxygen saturation during the first 24 hours of lifeArchives of Disease in Childhood: Fetal & Neonatal, 2000
- Individualised pulse oximetry limits in neonatal intensive careArchives of Disease in Childhood: Fetal & Neonatal, 1999
- Major reduction in alarm frequency with a new pulse oximeterIntensive Care Medicine, 1998
- Statistics from the inside. 14. Diagnostic and screening tests and reference values.Archives of Disease in Childhood, 1995
- Crying wolf: false alarms in a pediatric intensive care unit.1994
- Noninvasive Monitoring of Oxygenation in Infants and Children: Practical Considerations and Areas of ConcernPediatrics, 1994
- Die Meßgenauigkeit des Kontron 7840, Nellcor N200, und Radiometer OXI3 Pulsoximeters bei Säuglingen und KleinkindernKlinische Padiatrie, 1993
- Reliability of a pulse oximeter in the detection of hyperoxemiaThe Journal of Pediatrics, 1993
- A Cohort Study of Transcutaneous Oxygen Tension and the Incidence and Severity of Retinopathy of PrematurityNew England Journal of Medicine, 1992
- Hyperoxemia in Newborn Infants: Detection by Pulse OximetryPediatrics, 1989