Characterization of Low-Molecular-Weight Polymers: Failure of Universal Calibration in Size Exclusion Chromatography
- 1 April 1995
- journal article
- research article
- Published by Taylor & Francis in International Journal of Polymer Analysis and Characterization
- Vol. 1 (1) , 3-34
- https://doi.org/10.1080/10236669508009704
Abstract
Accurate measurement of molecular weights for polymers in the 500–5000 molecular weight range requires considerable care. Absolute methods such as membrane osmometry or light scattering are generally imprecise because of membrane permeation or low scattering intensity. Many polymers are too fragile to be studied by mass spectrometry. Other techniques, such as vapor phase osmometry, cryoscopy, and ebulliometry require calibration and data replication if accurate molecular weights are to be obtained. Size exclusion chromatography (SEC) also requires calibration but can be a very rapid and accurate method if proper calibration can be achieved. SEC calibration is now often achieved via “universal calibration” (UC). which is based on the premise that polymers with the same hydrodynamic volume will have the same SEC retention time, At very low molecular weights, the UC approach is clearly invalid since intrinsic viscosities can become negative for certain polymers. We demonstrate here that the invalidity of the UC concept extends well outside this molecular weight region, up to at least 1000 molecular weight for the polystyrene (PS)/polyisobutene (PIB) case. We have also tested whether PS. PIB. and n-alkanes elute at equivalent radii of gyration, Rg. This approach leads to somewhat smaller errors than UC, but also is not adequate for accurate work. Another SEC issue in this molecular weight range, with the commonly used differential refractive index (DRI) detector, is the variation of the detector response with molecular weight. This variation is proportional to 1/M n and. if neglected, can cause substantial overestimations of M n (e.g., 10–25%) in the SEC analysis of polydis-perse samples in THF solvent. Theoretical and experimental data are presented for PS, PIB. and polymethylmethacrylate, quantifying the error. Errors in light scattering weight-average molecular weights caused by the RI variation can also be significant and depend on the molecular weight distribution; accurate results can be obtained, but only if M n is accurately known, even if dn/dc is measured on the sample under study.Keywords
This publication has 31 references indexed in Scilit:
- Diffusion of oligomeric species in polymer solutionsMacromolecules, 1993
- Intrinsic viscosity of oligo- and polyisobutylenes: treatments of negative intrinsic viscositiesMacromolecules, 1991
- Determination of Polymer Molecular Weight by GPC in the Low Molecular Weight RegionJournal of Liquid Chromatography, 1990
- Viscosity and self-diffusion coefficient of linear polyethyleneMacromolecules, 1987
- Thermodynamic and hydrodynamic properties of the the systems polymer—tetrahydrofuran—water: 1. Solution properties of polystyrenePolymer, 1979
- Gel‐permeation chromatography of low molecular weight polymersJournal of Polymer Science Part C: Polymer Letters, 1976
- The nature of the double bond in low molecular weight polyisobutylenes and “polybutene” copolymersJournal of Polymer Science: Polymer Symposia, 1976
- Gel permeation chromatography: A guide to data interpretationPolymer Engineering & Science, 1973
- Refractive index increments of polymers in solution: 2. Refractive index increments and light-scattering in polydisperse systems of low molecular weightPolymer, 1972
- Variation of refractive index of polystyrene with molecular weight: Effect on the determination of molecular weight distributionsJournal of Applied Polymer Science, 1968