Abstract
A dry‐packing method has been developed which enables the preparation of packed capillary columns for micro HPLC from 250 μm i.d. tubing and 5 μm packing materials. Pressurized gases, such as hydrogen and argon, were used to transport the packing media, and ethanol or methanol were used as discharge agents. By changing the gas pressure, the packing density could be easily adjusted. It was found that, within experimental limits, the higher the packing density, the greater the column efficiency. Comparison between dry‐ and slurrypacked columns showed that the former had greatly improved stability; the efficiency of dry‐packed columns was about the same as, or even better than, that obtained by slurry‐packing.Keywords
This publication has 7 references indexed in Scilit:
- Inner wall coating of micro‐LC columns. Wall effects in LCJournal of High Resolution Chromatography, 1989
- Preparation and evaluation of dry-packed capillary columns for high-performance liquid chromatographyAnalytical Chemistry, 1988
- Quantitative analytical aspects of reversed-phase liquid chromatography with slurry-packed capillary columnsJournal of Chromatography A, 1987
- Dead-volume free termination for packed columns in microcapillary liquid chromatographyAnalytical Chemistry, 1984
- Performance evaluation of slurry-packed capillary columns for liquid chromatographyChromatographia, 1983
- Fused-silica narrow-bore microparticle-packed-column high-performance liquid chromatographyJournal of Chromatography A, 1982
- A study of micro-high-performance liquid chromatographyJournal of Chromatography A, 1977