Background— The purpose of this study was to determine the radiation exposure during catheter ablation of atrial fibrillation (AF) using the pulmonary vein (PV) approach. Methods and Results— The study included 15 patients with AF and 5 patients each with atrial flutter and atrioventricular nodal reentrant tachycardia (AVNRT) who underwent fluoroscopically guided procedures on a biplane x-ray system operated at a low-frame pulsed fluoroscopy (7.5 frames per second). Radiation exposure was measured directly with 50 to 60 thermoluminescent dosimeters (TLDs). Peak skin doses (PSDs), effective radiation doses, and risk of fatal malignancies were all computed. Mean fluoroscopy durations for AF procedures were 67.8±21 minutes in the right anterior oblique (RAO) and 61.9±16.6 minutes in the left anterior oblique (LAO) projection, significantly different from that required for atrial flutter and AVNRT. The mean PSDs measured with the TLDs were 1.0±0.5 Gy in the RAO and 1.5±0.4 Gy in the LAO projection. The lifeti...Keywords
This publication has 12 references indexed in Scilit:
- Predictors of Recurrence Following Catheter Ablation of Atrial Fibrillation Using an Irrigated‐Tip Ablation CatheterJournal of Cardiovascular Electrophysiology, 2004
- Technique and Results of Pulmonary Vein Angiography in Patients Undergoing Catheter Ablation of Atrial FibrillationJournal of Cardiovascular Electrophysiology, 2004
- Radiation Exposure During Radiofrequency Catheter Ablation for Atrial FibrillationPacing and Clinical Electrophysiology, 2003
- Accurate Assessment of Patient Effective Radiation Dose and Associated Detriment Risk From Radiofrequency Catheter Ablation ProceduresCirculation, 2001
- Fluoroscopy: Patient Radiation Exposure IssuesRadioGraphics, 2001
- Spontaneous Initiation of Atrial Fibrillation by Ectopic Beats Originating in the Pulmonary VeinsNew England Journal of Medicine, 1998
- Predictors of fluoroscopy time and estimated radiation exposure during radiofrequency catheter ablation proceduresThe American Journal of Cardiology, 1998
- Acute Radiation Dermatitis Following Radiofrequency Catheter Ablation of Atrioventricular Nodal Reentrant TachycardiaPacing and Clinical Electrophysiology, 1997
- Fluoroscopy and the skin: Implications for radiofrequency catheter ablationThe American Journal of Cardiology, 1995
- Radiation exposure to patients and medical personnel during radiofrequency catheter ablation for supraventricular tachycardiaThe American Journal of Cardiology, 1992