Online coupling of in vivo microdialysis with tandem mass spectrometry

Abstract
The capability of interfacing in vivo microdialysis with mass spectrometry has been demonstrated. The goal of this research was to demonstrate the feasibility of real-time analysis in biological systems using microdialysis in combination with tandem mass spectrometry (MS/MS). Microdialysis sampling was accomplished by surgically implanting a small microdialysis probe into a tissue or area of interest. Molecules diffuse through the membrane of the microdialysis probe due to concentration differences. These molecules are collected in a sample loop and analyzed by tandem mass spectrometry. Sequential injections can be made in as little as 2 min. This capability is advantageous in the study of molecules with very rapid elimination rates. Tris(2-chloroethyl)phosphate (TRCP) was used as a model compound in the development of this analytical technique. As an example of an application of the microdialysis/MS/MS technique, plasma concentration vs time curves were obtained and compared with the plasma concentration profiles obtained using conventional studies. For the microdialysis/MS/MS studies, the average slope from three animals was -0.086 min-1. In comparison, the average slope from four animals from the conventional studies was -0.035 min-1.