A 2-dimensional detector decoding study on BGO arrays with quadrant sharing photomultipliers
- 1 August 1994
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 41 (4) , 1453-1457
- https://doi.org/10.1109/23.322918
Abstract
This is a theoretical estimation and experimental study of the positron camera detector design using quadrant sharing of phototubes and optically cross-coupled BGO scintillation crystals. This study estimates the maximum number of BGO crystals which can be coupled to a photomultiplier and be decoded with this design, based on the number of photoelectrons generated and the optical efficiency of the crystal-phototube combination. From photoelectron statistics, a 10×10 BGO array can be coupled to a single square phototube (Hamamatsu R-2497) and have each crystal clearly decoded. Simulation of 1-dimensional and 2-dimensional statistical photoelectron distribution verifies the estimations. The simulation also yields the optimal distribution of light to the decoding phototubes. This theoretically optimal light distribution is used to guide the development of an 8×8 array of 3.2×3.2 mm BGO crystals (10 mm deep) using Hamamatsu R-2497 phototubes. All 64 crystals in the array can be clearly identified. The reconstructed image resolution is further estimated for the 8×8 and 10×10 arrays. These resolution results are also compared to discrete detectors which do not use analog decoding. The comparison shows that the analog decoding process degrades the image resolution by 7 to 9%Keywords
This publication has 8 references indexed in Scilit:
- Positioning and energy response of PET block detectors with different light sharing schemesIEEE Transactions on Nuclear Science, 1994
- Measurement of the coincidence response of very thin BGO crystalsIEEE Transactions on Nuclear Science, 1994
- A positron camera detector design with cross-coupled scintillators and quadrant sharing photomultipliersIEEE Transactions on Nuclear Science, 1993
- An improved multicrystal 2-D BGO detector for PETIEEE Transactions on Nuclear Science, 1992
- Detector, shielding and geometric design factors for a high-resolution PET systemIEEE Transactions on Nuclear Science, 1990
- An evaluation of a two-dimensional array detector for high resolution PETIEEE Transactions on Medical Imaging, 1988
- Initial Results from the Donner 600 Crystal Positron TomographIEEE Transactions on Nuclear Science, 1987
- A Multicrystal Two Dimensional BGO Detector System for Positron Emission TomographyIEEE Transactions on Nuclear Science, 1986