Calculation models for determining the absorbed dose in water phantoms in off-axis planes of rectangular fields of open and wedged photon beams
- 1 April 1995
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 40 (4) , 511-527
- https://doi.org/10.1088/0031-9155/40/4/003
Abstract
Beam models are proposed for the calculation of the dose in off-axis planes of rectangular photon fields, when the data set used in the treatment planning system is based on the simple storage model of Milan and Bentley (1974). For open beams the model separates the off-axis ratio into an envelope profile and two boundary profiles. The envelope profile gives the field intensity of the maximal position of the jaws and has rotational symmetry. The boundary profiles describe the boundaries of the field actually formed by the jaws. In the case of a wedged beam, the model also separates the off-axis ratio into envelope profiles and boundary profiles. To determine these profiles for the non-wedge direction from open beam profiles, the wedge thickness is converted to an equivalent water thickness. In the case of an asymmetric field, the boundary profiles are shifted to the field centre. Results of calculation with these models have been compared with measurements and the simple multiplication of profiles, which has often been used with the Milan-Bentley model. The new models agree within a few per cent with the measurements and are a great improvement compared to the simple multiplication of profiles.Keywords
This publication has 9 references indexed in Scilit:
- Relative output factors of asymmetric megavoltage beamsRadiotherapy and Oncology, 1994
- Comprehensive QA for radiation oncology: Report of AAPM Radiation Therapy Committee Task Group 40Medical Physics, 1994
- Monte Carlo calculations of scatter to primary ratios for normalisation of primary and scatter dosePhysics in Medicine & Biology, 1990
- Dose computations for asymmetric fields defined by independent jawsMedical Physics, 1988
- Dosimetry of asymmetric x‐ray collimatorsMedical Physics, 1986
- Off-center ratios for three-dimensional dose calculationsMedical Physics, 1986
- Errors in off-axis treatment planning for a 4-MeV machineMedical Physics, 1983
- The storage and manipulation of radiation dose data in a small digital computerThe British Journal of Radiology, 1974
- CorrespondenceThe British Journal of Radiology, 1966