Laparoscopic implant instrument for the placement of intramuscular electrodes in the diaphragm
- 1 September 1999
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Rehabilitation Engineering
- Vol. 7 (3) , 360-371
- https://doi.org/10.1109/86.788472
Abstract
We have developed an endoscopic instrument that will allow a surgeon to safely, dependably and accurately place intramuscular (IM) electrodes in the diaphragm. This instrument has been used to implant 28 IM electrodes in the diaphragms of eleven acute and four chronic dogs. All electrodes achieved full activation of the diaphragm muscle, producing tidal volumes up to 130% V/sub TCRIT/, the critical volume necessary for basal ventilatory support, with unilateral stimulation. The surgeon is able to control the angle of the IM electrode insertion needle, which enables the needle to approach the diaphragm at an angle that is parallel to the surface of the muscle. This insures good control over the depth of needle penetration into the muscle, which greatly reduces the risk of accidentally passing the needle through the diaphragm and entering the thorax. Endoscopic placement of IM electrodes into the diaphragm opens opportunities to provide cost effective negative pressure ventilation to patients who are unable to effect sufficient ventilation by central nervous system (CNS) control of respiration.Keywords
This publication has 22 references indexed in Scilit:
- The tissue response to epimysial electrodes for diaphragm pacing in dogsIEEE Transactions on Biomedical Engineering, 1997
- Tracheostomy ventilation versus diaphragmatic pacemaker ventilation in high spinal cord injurySpinal Cord, 1994
- 200 Cases with a New Breathing Pacemaker Dispel Myths about Diaphragm PacingAsaio Journal, 1994
- Long-term intramuscular electrical activation of the phrenic nerve: efficacy as a ventilatory prosthesisIEEE Transactions on Biomedical Engineering, 1994
- Long-term intramuscular electrical activation of the phrenic nerve: safety and reliabilityIEEE Transactions on Biomedical Engineering, 1994
- Diaphragmatic Pacing: An Option for Patients With Quadriplegic Respiratory ParalysisAORN Journal, 1993
- Electrical Activation of Respiratory Muscles by Methods Other than Phrenic Nerve Cuff ElectrodesPacing and Clinical Electrophysiology, 1989
- Comparative study of electrophrenic nerve stimulation and mechanical ventilatory support in traumatic spinal cord injurySpinal Cord, 1987
- Electrophrenic RespirationScience, 1948
- Production of Artificial Respiration by Rhythmic Stimulation of the Phrenic NervesNature, 1937