Rapid DNA Amplification Using a Battery-Powered Thin-Film Resistive Thermocycler
- 1 January 2009
- book chapter
- Published by Springer Nature
- Vol. 504, 441-458
- https://doi.org/10.1007/978-1-60327-569-9_24
Abstract
A prototype handheld, compact, rapid thermocycler was developed for multiplex analysis of nucleic acids in an inexpensive, portable configuration. Instead of the commonly used Peltier heating/cooling element, electric thin-film resistive heater and a miniature fan enable rapid heating and cooling of glass capillaries leading to a simple, low-cost Thin-Film Resistive Thermocycler (TFRT). Computer-based pulse width modulation control yields heating rates of 6–7 K/s and cooling rates of 5 K/s. The four capillaries are closely coupled to the heater, resulting in low power consumption. The energy required by a nominal PCR cycle (20 s at each temperature) was found to be 57 ± 2 J yielding an average power of approximately 1.0 W (not including the computer and the control system). Thus the device can be powered by a standard 9 V alkaline battery (or other 9 V power supply). The prototype TFRT was demonstrated (in a benchtop configuration) for detection of three important food pathogens (E. coli ETEC, Shigella dysenteriae, and Salmonella enterica). PCR amplicons were analyzed by gel electrophoresis. The 35 cycle PCR protocol using a single channel was completed in less then 18 min. Simple and efficient heating/cooling, low cost, rapid amplification, and low power consumption make the device suitable for portable DNA amplification applications including clinical point of care diagnostics and field use.Keywords
This publication has 49 references indexed in Scilit:
- A “do-it-yourself” array biosensorMethods, 2005
- Genotyping of Clostridium perfringens toxins using multiple oligonucleotide microarray hybridizationMolecular and Cellular Probes, 2004
- On-Chip Nanoliter-Volume Multiplex TaqMan Polymerase Chain Reaction from A Single Copy Based on Counting Fluorescence Released MicrochambersAnalytical Chemistry, 2004
- Identification of Bacillus anthracis by multiprobe microarray hybridizationDiagnostic Microbiology and Infectious Disease, 2004
- Capillary Electrophoresis Chips with a Sheath-Flow Supported Electrochemical Detection SystemAnalytical Chemistry, 2004
- Multiple sequence alignment with the Clustal series of programsNucleic Acids Research, 2003
- Molecular Beacons: Probes that Fluoresce upon HybridizationNature Biotechnology, 1996
- A one-step coupled amplification and oligonucleotide ligation procedure for multiplex genetic typing.Genome Research, 1995
- CLUSTAL W: improving the sensitivity of progressive multiple sequence alignment through sequence weighting, position-specific gap penalties and weight matrix choiceNucleic Acids Research, 1994
- An inexpensive and versatile computer-controlled PCR machine using a Peltier element as thermoelectric heat pumpTrends in Genetics, 1989