Dosimetric properties of the Theraview fluoroscopic electronic portal imaging device.
- 1 May 2000
- journal article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 73 (869) , 517-530
- https://doi.org/10.1259/bjr.73.869.10884749
Abstract
Electronic portal imaging devices (EPIDs) can be used for non-imaging applications in radiotherapy such as patient dosimetry. Of the systems available, the fluoroscopic camera-based EPID Theraview (InfiMed Inc.) has not been studied to date, and a review of the dosimetric properties of the system is presented here. In the "single set-up" mode of image acquisition, pixel intensity increases sublinearly with applied dose. The response was dependent on the system's video signal gain and showed a threshold dose to the detector in the range 0.05-0.35 cGy, and pixel saturation at detector doses in the range 1.2-1.6 cGy. Repeated exposures of the EPID were observed to be extremely reproducible (standard deviation 0.5%). The sensitivity of the system showed a linear decline of 0.04% day-1 over a 68-day period, during which time the relative off-axis response within 10 x 10 cm2 field was constant to within a standard deviation of 0.56%. The system shows spatial non-uniformity, which requires correction for application to dose measurements in two-dimensions. Warm-up of the camera control unit required a period of at least 40 min and was associated with an enhancement in pixel intensity of up to 12%. A radiation dose history effect was observed at doses as low as 0.2 Gy. Camera dark current was shown to be negligible at normal accelerator operation. No discernible image distortion was found. Mechanical stability on gantry rotation was also assessed and image displacement of up to 5 mm at the isocentre was observed. It was concluded that the device could be used for dosimetry provided necessary precautions were observed and corrections made.Keywords
This publication has 34 references indexed in Scilit:
- Assessing radiation and light field congruence with a video based electronic portal imaging deviceMedical Physics, 1996
- Evaluation of electronic portal imaging device for missing tissue compensator design and verificationMedical Physics, 1995
- Portal dose measurement in radiotherapy using an electronic portal imaging device (EPID)Physics in Medicine & Biology, 1995
- Daily dosimetric quality control of the MM50 Racetrack Microtron using an electronic portal imaging deviceRadiotherapy and Oncology, 1995
- Portal imaging technology: Past, present, and futureSeminars in Radiation Oncology, 1995
- A multipurpose phantom for use with electronic portal imaging devicesPhysics in Medicine & Biology, 1995
- A technique of automating compensator design for lung inhomogeneity correction using an electron portal imaging deviceMedical Physics, 1994
- The lens coupling efficiency in megavoltage imagingMedical Physics, 1991
- Performance of a prototype fluoroscopic radiotherapy imaging systemInternational Journal of Radiation Oncology*Biology*Physics, 1990