Low-level internal radionuclide contamination: use of gamma camera for detection.
- 1 January 1984
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 150 (1) , 235-240
- https://doi.org/10.1148/radiology.150.1.6689766
Abstract
The uncollimated .gamma. camera was evaluated as a screening tool for the detection of low levels of internal radionuclide contamination [in humans]. The minimum detectable activity of (MDA) is on the order of 1-4 nCi (37-148 Bq [becquerels]) for a 3-min sample and background counting time. Identification of unknown radionuclides at these low levels involves a systematic search and hence requires a more careful peaking method than organ imaging. Most .gamma. cameras are capable both of detecting a wide range of (35 keV to 1.332 MeV) of .gamma.-ray energies in the presence of 3-4 radionuclides and also of distinguishing principal .gamma.-ray energies from Compton scatter. The differences in MDA among .gamma. cameras are small, and vary primarily with the background activity. The actual contaminant activities lie between the calculated values based on measurements when the source is 10 cm away from the uncollimated detector head and using point sources both in air and with the tissue-equivalent scatter medium. The widely available .gamma. camera is capable of screening a large population, either suspected or actually contaminated, in the event of an emergency.This publication has 4 references indexed in Scilit:
- Exposure to Xenon 133 in the nuclear medicine laboratory.Radiology, 1982
- DEVELOPMENT OF A CORRECTION FACTOR FOR XE-133 VIALS FOR USE WITH A DOSE CALIBRATOR1982
- The Effectiveness of a Whole Body Counter During and After an Accident Situation at Nuclear FacilitiesHealth Physics, 1981
- The impact of a nuclear crisis on a radiology department.Radiology, 1980