Use of recombinant human cardiac Troponin T for standardization of third generation Troponin T methods
- 1 January 1999
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 59 (sup230) , 128-131
- https://doi.org/10.1080/00365519909168336
Abstract
The Elecsys® Troponin T third generation assay uses recombinant human cardiac troponin T as standard material. The new assay has a linear calibration curve. Thus, linearity problems observed with the second generation assay have been eliminated. The assay has high precision, especially at the low end of measuring range (inter-assay CV <10 % at 0.1 μg/L). The new standardization does not change the cut-off value of 0.1 μg/L. The use of recombinant human cardiac troponin T as standard material enables a reproducible and reliable standardization of troponin T assays.Keywords
This publication has 1 reference indexed in Scilit: