Using a triple‐quadrupole mass spectrometer in accurate mass mode and an ion correlation program to identify compounds
Open Access
- 26 August 2005
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 19 (18) , 2699-2715
- https://doi.org/10.1002/rcm.2112
Abstract
Atomic masses and isotopic abundances are independent and complementary properties for discriminating among ion compositions. The number of possible ion compositions is greatly reduced by accurately measuring exact masses of monoisotopic ions and the relative isotopic abundances (RIAs) of the ions greater in mass by +1 Da and +2 Da. When both properties are measured, a mass error limit of 6–10 mDa (< 31 ppm at 320 Da) and an RIA error limit of 10% are generally adequate for determining unique ion compositions for precursor and fragment ions produced from small molecules (less than 320 Da in this study). ‘Inherent interferences’, i.e., mass peaks seen in the product ion mass spectrum of the monoisotopic [M+H]+ ion of an analyte that are −2, −1, +1, or +2 Da different in mass from monoisotopic fragment ion masses, distort measured RIAs. This problem is overcome using an ion correlation program to compare the numbers of atoms of each element in a precursor ion to the sum of those in each fragment ion and its corresponding neutral loss. Synergy occurs when accurate measurement of only one pair of +1 Da and +2 Da RIAs for the precursor ion or a fragment ion rejects all but one possible ion composition for that ion, thereby indirectly rejecting all but one fragment ion-neutral loss combination for other exact masses. A triple-quadrupole mass spectrometer with accurate mass capability, using atmospheric pressure chemical ionization (APCI), was used to measure masses and RIAs of precursor and fragment ions. Nine chemicals were investigated as simulated unknowns. Mass accuracy and RIA accuracy were sufficient to determine unique compositions for all precursor ions and all but two of 40 fragment ions, and the two corresponding neutral losses. Interrogation of the chemical literature provided between one and three possible compounds for each of the nine analytes. This approach for identifying compounds compensates for the lack of commercial ESI and APCI mass spectral libraries, which precludes making tentative identifications based on spectral matches. Published in 2005 by John Wiley & Sons, Ltd.Keywords
This publication has 18 references indexed in Scilit:
- Non-regulated water contaminants: emerging researchEnvironmental Impact Assessment Review, 2004
- Mass spectrometric identification of an azobenzene derivative produced by smectite-catalyzed conversion of 3-amino-4-hydroxyphenylarsonic acidTalanta, 2003
- Utility of three types of mass spectrometers for determining elemental compositions of ions formed from chromatographically separated compoundsRapid Communications in Mass Spectrometry, 2002
- Exact Mass Measurements for Confirmation of Pesticides and Herbicides Determined by Liquid Chromatography/Time-of-Flight Mass SpectrometryAnalytical Chemistry, 2001
- Screening and identification of unknown contaminants in water with liquid chromatography and quadrupole-orthogonal acceleration-time-of-flight tandem mass spectrometryJournal of Chromatography A, 2001
- Pharmaceuticals and Personal Care Products in the Waters of Lake Mead, NevadaPublished by American Chemical Society (ACS) ,2001
- Ion Composition Elucidation (ICE): an Investigative Tool for Characterization and Identification of Compounds of Regulatory ImportanceEnvironmental Forensics, 2001
- A mass peak profile generation model to facilitate determination of elemental compositions of ions based on exact masses and isotopic abundancesJournal of the American Society for Mass Spectrometry, 1997
- Determination of Elemental Compositions from Mass Peak Profiles of the Molecular Ion (M) and the M + 1 and M + 2 IonsAnalytical Chemistry, 1996
- Mass Measurements by an Accurate and Sensitive Selected Ion Recording TechniqueAnalytical Chemistry, 1994