How Secure Are Laparoscopically Placed Clips?
- 1 June 1992
- journal article
- research article
- Published by American Medical Association (AMA) in Archives of Surgery
- Vol. 127 (6) , 718-720
- https://doi.org/10.1001/archsurg.1992.01420060094014
Abstract
• We tested the security of laparoscopic, hemostatic clips in vitro and in vivo. In vitro, the force needed to transversely dislodge clips that were applied to silicone (Silastic) tubing ranged from a mean±SEM of 1.81 ±0.07 N (Endo Clip) to 2.68±0.04 N (Ligaclip). The force needed to axially dislodge clips that were applied to silicone tubing ranged from 4.08±0.20 N (Endo Clip) to 4.78±0.18 N (Ligaclip). In vivo, on mesenteric vessels in anesthetized pigs, 34 of 100 clips that were applied laparoscopically with the Endo Clip were moved or dislodged compared with 11 of 100 of those applied with the Ligaclip. We conclude that laparoscopically placed vascular clips applied with the Ligaclip are more secure than those applied with the Endo Clip. The laparoscopic surgeon should be aware of the relative ease with which clips may be dislodged. (Arch Surg. 1992;127:718-720)Keywords
This publication has 3 references indexed in Scilit:
- Ligation of the structures of the cystic pedicle during laparoscopic cholecystectomyThe American Journal of Surgery, 1991
- Minimal access general surgery: the dawn of a new eraThe American Journal of Surgery, 1991
- Laparoscopic guided cholecystectomyThe American Journal of Surgery, 1991