Electrocautery-induced pacemaker malfunction during surgery
- 1 July 1991
- journal article
- case report
- Published by Springer Nature in Canadian Journal of Anesthesia/Journal canadien d'anesthésie
- Vol. 38 (5) , 616-618
- https://doi.org/10.1007/bf03008198
Abstract
We report a case of electrocautery-induced pacemaker failure that resulted in asystole in a 15-year-old girl scheduled for cardiac surgery. Her pacemaker was converted to the asynchronous mode the night before surgery. Electromagnetic interference from the unipolar electrocautery caused a reduction in the battery voltage, which allowed the digital circuitry, but not the voltage control oscillator (VCO), to work properly. Eventually the battery current drain caused VCO "lock-out," and pacemaker and battery failure. This report demonstrates that electrocautery-induced pacemaker failure can occur, even after conversion to asynchronous mode.Keywords
This publication has 9 references indexed in Scilit:
- SAFETY PRECAUTIONS IN THE MANAGEMENT OF PATIENTS WITH PACEMAKERS WHEN ELECTROCAUTERY OPERATIONS ARE PERFORMED1988
- Electrocautery and Pacemakers: Management of the Paced Patient Subject to ElectrocauteryThe Annals of Thoracic Surgery, 1986
- PacemakersAnesthesiology, 1984
- Effect of coagulating and cutting current on a demand pacemaker during transurethral resection of the prostate. a case reportCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1978
- Perioperative Management of the Pacemaker PatientAnesthesiology, 1977
- Anesthesia for Patients with Heart Block and Artificial Cardiac PacemakersAnesthesia & Analgesia, 1976
- Suppression of a Demand Pacemaker by Transurethral ElectrocauteryAnesthesia & Analgesia, 1973
- Pacemaker Malfunction From Urethral ElectrocauteryJAMA, 1971
- Cardiac pacemakers as an anaesthetic problemAnaesthesia, 1970