X-ray flash techniques
- 1 May 1979
- journal article
- Published by IOP Publishing in Journal of Physics E: Scientific Instruments
- Vol. 12 (5) , 336-350
- https://doi.org/10.1088/0022-3735/12/5/001
Abstract
The author examines the problems of X-ray flash production and its applications, similar to those of continuous X-radiation. Special problems are the fast production of high X-ray intensities to produce detectable radiation and the high-voltage and high-current discharge circuits required. Anode voltages from some 10 kV to about 1 MV and currents up to 10 kA are normal for laboratory equipment, but 100 MV and 1 MA are possible. The flash time is mostly between 10 ns and 1 mu s. Cineradiography with frame rates up to 106 pictures s-1 can be achieved.Keywords
This publication has 30 references indexed in Scilit:
- Some aspects of high-current relativistic electron-beam generationJournal of Vacuum Science and Technology, 1975
- Investigation of micro-channel plates as parallel-bore electron collimators for use in a proximity-focused ultra-fast streak tubeNuclear Instruments and Methods, 1975
- High resolution position-sensitive detectors using microchannel platesNuclear Instruments and Methods, 1974
- A 400 rad flash x-ray system for pulse radiolysisReview of Scientific Instruments, 1974
- The magnetic pressure in cylindrical wires and rectangular rods, and the inverse pinch effectJournal of Physics D: Applied Physics, 1973
- Some Characteristics of a 600 kV Flash X-ray TubePhysica Scripta, 1972
- Étude de l'amorçage des étincelles sous vide à l'aide d'une étincelle auxiliaire produite par une impulsion de courant rectangulaireRevue de Physique Appliquée, 1970
- Fluorescent Screens for Monochromatic X-Ray Diffraction Patterns on Polaroid FilmReview of Scientific Instruments, 1968
- Oregon SectionAmerican Journal of Physics, 1965
- Experiments with a Condenser Discharge X-Ray TubePhysical Review B, 1938