A noninvasive estimation of mixed venous oxygen saturation using near‐infrared spectroscopy by cerebral oximetry in pediatric cardiac surgery patients
- 23 May 2005
- journal article
- research article
- Published by Wiley in Pediatric Anesthesia
- Vol. 15 (6) , 495-503
- https://doi.org/10.1111/j.1460-9592.2005.01488.x
Abstract
Near-infrared spectroscopy (NIRS) is a noninvasive optical monitor of regional cerebral oxygen saturation (rSO2). The aim of this study was to validate the use of NIRS by cerebral oximetry in estimating invasively measured mixed venous oxygen saturation (SvO2) in pediatric postoperative cardiac surgery patients. Twenty patients were enrolled following cardiac surgery with intraoperative placement of a pulmonary artery (PA) or superior vena cava (SVC) catheter. Five patients underwent complete biventricular repair--complete atrioventricular canal (n=3) and other (n=2). Fifteen patients with functional single ventricle underwent palliative procedures--bidirectional Glenn (n=11) and Fontan (n=4). Cerebral rSO2 was monitored via NIRS (INVOS 5100) during cardiac surgery and 6 h postoperatively. SvO2 was measured from blood samples obtained via an indwelling PA or SVC catheter and simultaneously correlated with rSO2 by NIRS at five time periods: in the operating room after weaning from cardiopulmonary bypass, after sternal closure, and in the CICU at 2, 4, and 6 h after admission. Each patient had five measurements (total=100 comparisons). SvO2 obtained via an indwelling PA or SVC catheter for all patients correlated with rSO2 obtained via NIRS: Pearson's correlation coefficient of 0.67 (P<0.0001) and linear regression of r2=0.45 (P<0.0001). Separate linear regression of the complete biventricular repairs demonstrated an r=0.71, r2=0.50 (P<0.0001). Bland-Altman analysis showed a bias of +3.3% with a precision of 16.6% for rSO2 as a predictor of SvO2 for all patients. Cerebral rSO2 was a more accurate predictor of SvO2 in the biventricular repair patients (bias -0.3, precision 11.8%), compared with the bidirectional Glenn and Fontan patients. Regional cerebral oximetry via NIRS correlates with SvO2 obtained via invasive monitoring. However, the wide limits of agreement suggest that it may not be possible to predict absolute values of SvO2 for any given patient based solely on the noninvasive measurement of rSO2. Near-infrared spectroscopy, using the INVOS 5100 cerebral oximeter, could potentially be used to indicate trends in SVO2, but more studies needs to be performed under varying clinical conditions.Keywords
This publication has 29 references indexed in Scilit:
- Predictive Factors for Usefulness of Fiberoptic Pulmonary Artery Catheter for Continuous Oxygen Saturation in Mixed Venous Blood Monitoring in Cardiac SurgeryAnesthesia & Analgesia, 1997
- Cerebral oxygenation during cardiopulmonary bypass in childrenThe Journal of Thoracic and Cardiovascular Surgery, 1997
- Central venous thrombosis after cardiac operations in childrenThe Journal of Thoracic and Cardiovascular Surgery, 1996
- Usefulness of intermittent monitoring of mixed venous oxygen saturation after stage I palliation for hypoplastic left heart syndromeThe American Journal of Cardiology, 1994
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Complications of vascular catheterization in critically ill childrenCritical Care Medicine, 1984
- Noninvasive, Infrared Monitoring of Cerebral and Myocardial Oxygen Sufficiency and Circulatory ParametersScience, 1977
- On-line monitoring of mixed venous oxygen saturation after cardiothoracic surgery.Thorax, 1975
- Critical Assessment of Use of Central Venous Oxygen Saturation as a Mirror of Mixed Venous Oxygen in Severely Ill Cardiac PatientsCirculation, 1969
- Estimation of Cardiac Output Soon After Intracardiac Surgery with Cardiopulmonary BypassAnnals of Surgery, 1959