Duplex Doppler sonography of the carotid artery: false-positive results in an artery contralateral to an artery with marked stenosis.

Abstract
Anecdotal reports have described a false-positive "jet effect" or velocity increase in the carotid artery contralateral to an artery with significant stenosis or occlusion when using duplex Doppler sonography. In this study, the frequency, significance, and possible reasons for this finding were evaluated by a retrospective comparison of duplex sonography and angiography. Twenty-three patients with unilateral 81-100% carotid artery stenosis who underwent both duplex sonography (16 Acuson, seven Quantum) and angiography were evaluated. In 14 patients, there was an accurate or slight underestimate (less than 20%) of stenosis present in the internal carotid artery contralateral to an artery with tight stenosis/occlusion. In nine, a velocity increase in the internal carotid artery resulted in overestimation (10-80%) of the actual degree of stenosis. In one of these nine patients, real-time images were sufficient to explain the velocity increase on the basis of vessel tortuosity. In one, falsely elevated veloc...