Determination of 25-hydroxyvitamin D3 in human serum by fluorescence labelling and high-performance liquid chromatography
- 1 January 1991
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in The Analyst
- Vol. 116 (12) , 1347-1351
- https://doi.org/10.1039/an9911601347
Abstract
A method is described for the determination of 25-hydroxyvitamin D3 in human blood serum. The problems of sensitivity and selectivity encountered with previous techniques were avoided by the formation of a highly fluorescent Diels–Alder adduct following solid-phase extraction of the vitamin. After excess of reagent had been eliminated, quantification was achieved by high-performance liquid chromatography. The recovery of the vitamin from serum was 76.4 ± 1.76%. The precision of the method was determined, and the relative standard deviations were 8.38% at a concentration of 47.0 × 10–9 mol dm–3, 6.74% at a concentration of 99.8 × 10–9 mol dm–3 and 3.79% at a concentration of 146.8 × 10–9 mol dm–3. The detection limit for the adduct was 2.93 × 10–14 mol injected, for a signal-to-noise ratio of 3 : 1, and serum concentrations of 0.25 × 10–9 mol dm–3 could easily be quantified. No interference from endogenous or exogenous substances was observed.Keywords
This publication has 0 references indexed in Scilit: