Determination of aluminium in dental enamel by the carbon cup atomic-absorption method
- 1 January 1975
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in The Analyst
- Vol. 100 (1197) , 884-890
- https://doi.org/10.1039/an9750000884
Abstract
The possibility of using the carbon cup atomisation technique for the atomic-absorption determination of aluminium in dental enamel has been investigated. Various parameters that influence the sensitivity and accuracy of the measurements were studied by use of a standard electronic detection system (Varian AA-5; chart recorder f.s.d., 1 s) and oscilloscopic (Tektronix) recording of the absorption signal. The latter system enabled the atomic-and non-specific absorption signals to be studied as a function of the time of atomisation. Thus, increasing the rate of heating of the carbon cup in the atomisation step resulted in a substantial enhancement of the aluminium peak absorption signal. The presence of calcium in the solution (7·2 mg ml–1, in the form of the phosphate) produced a two-fold increase in sensitivity for aluminium. For routine determinations of aluminium in tooth material (in vitro) 10 mg of finely ground and homogenised sample are dissolved in 6 N nitric acid and the solution is diluted to 0·5 ml; 5-µl aliquots are sampled for a single determination. A single calibration graph obtained by the standard addition method is adequate for one set of samples. No correction due to non-specific absorption is necessary, but the blank value (nitric acid) should be checked from time to time. The absolute sensitivity of the method (3·5 × 10–11 g of aluminium) permits the determination of aluminium in tooth material both in vitro and in vivo at microgram per gram concentration levels with a general precision of 8 per cent. In vivo analysis can be carried out on 300–600 µg of sample obtained by a chemical biopsy method. The main problem encountered in in vivo analysis lies in obtaining an aluminium-free support material for etching purposes. In addition, “clean room” facilities are recommended for improving the accuracy.Keywords
This publication has 0 references indexed in Scilit: