A Practical Ion Trap Mass Spectrometer for the Analysis of Peptides by Matrix-Assisted Laser Desorption/Ionization
- 1 January 1996
- journal article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 68 (10) , 1784-1791
- https://doi.org/10.1021/ac9511612
Abstract
The present paper describes the performance of a newly configured matrix-assisted laser desorption/ionization quadrupole ion trap mass spectrometer (MALDI-ITMS), designed for biological applications that require the determination of the primary structures of proteins, e.g., the rapid identification of proteins and the elucidation of posttranslational modifications. The strategy used for solving problems of this type involves enzymatic digestion of the protein, followed by MALDI ion trap mass spectrometric analysis of the components of the resulting complex mixture of peptide ions. The new instrument is demonstrated to be a highly practical tool for analyzing proteins. In particular, mixtures containing as many as 30 peptide components can be rapidly and sensitively analyzed without prior chromatographic separation of the components. Informative tandem mass spectra can be obtained from the peptide components with m/z values up to 3500. A single subpicomole sample loading of a complex peptide mixture is more than sufficient for a complete set of experiments that includes both low- and high-resolution molecular mass determinations as well as a complete MS/MS study of the various components present in the sample. Extensive use is made of improved methods for trapping, isolating, fragmenting, and detecting ions in the ITMS (details are to be provided in two future papers).Keywords
This publication has 12 references indexed in Scilit:
- Matrix Dependence of Metastable Fragmentation of Glycoproteins in MALDI TOF Mass SpectrometryAnalytical Chemistry, 1995
- Reduction of space‐charging in the quadrupole ion trap by sequential injection and simultaneous storage of positively and negatively charged ionsRapid Communications in Mass Spectrometry, 1993
- Matrix‐assisted laser desorption of peptides and proteins on a quadrupole ion trap mass spectrometerRapid Communications in Mass Spectrometry, 1993
- Matrix-assisted laser desorption/ionization inside a quadrupole ion-trap detector cellRapid Communications in Mass Spectrometry, 1992
- Improved accuracy of mass measurement with a quadrupole ion‐trap mass spectrometerRapid Communications in Mass Spectrometry, 1992
- Matrix-Assisted Laser Desorption/Ionization Mass Spectrometry of BiopolymersAnalytical Chemistry, 1991
- Operation of a quadrupole ion trap mass spectrometer to achieve high mass/charge ratiosInternational Journal of Mass Spectrometry and Ion Processes, 1991
- High resolution on a quadrupole ion trap mass spectrometerJournal of the American Society for Mass Spectrometry, 1991
- Ion isolation and sequential stages of mass spectrometry in a quadrupole ion trap mass spectrometerInternational Journal of Mass Spectrometry and Ion Processes, 1990
- Instrumentation, applications, and energy deposition in quadrupole ion-trap tandem mass spectrometryAnalytical Chemistry, 1987