Application of a video-optical beam imaging system for quality assurance of medical accelerators

Abstract
Method validation techniques were developed and experiments were carried out using a beam imaging system (BIS, Wellhöfer Dosimetrie, Schwarzenbruck, Germany) for routine quality assurance of medical accelerators. The routine quality assurance tasks include x-ray beam flatness and symmetry check, light/radiation field congruence test, beam energy constancy for electrons and mechanical checks for couch and collimator rotations. Comparisons were made between the BIS application and conventional quality assurance methods that use radiographic films or detector arrays. In this work, we have demonstrated efficiency and accuracy of the BIS to perform some of the routine quality assurance tasks for medical linear accelerators.