Thin film microcalorimeter for heat capacity measurements from 1.5 to 800 K
- 1 April 1994
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 65 (4) , 946-959
- https://doi.org/10.1063/1.1144925
Abstract
A new microcalorimeter for measuring heat capacity of thin films in the range 1.5–800 K is described. Semiconductor processing techniques are used to create a device with an amorphous silicon nitride membrane as the sample substrate, a Pt thin film resistor for temperatures greater than 40 K, and either a thin film amorphous Nb–Si or a novel boron-doped polycrystalline silicon thermometer for lower temperatures. The addenda of the device, including substrate, is 4×10−6 J/K at room temperature and 2×10−9 J/K at 4.3 K, approximately two orders of magnitude less than any existing calorimeter used for measuring thin films. The device is capable of measuring the heat capacity of thin film samples as small as a few micrograms.Keywords
This publication has 25 references indexed in Scilit:
- Heat capacity and sticking probability measurements ofsubmonolayers adsorbed on evaporated Ag films: Bose statistics in two dimensionsPhysical Review Letters, 1990
- On the Temperature Dependence of Resistivity of Polycrystalline Silicon FilmsMRS Proceedings, 1987
- Tunneling and Transport Measurements at the Metal-Insulator Transition of Amorphous Nb: SiPhysical Review Letters, 1983
- Measurement of low-temperature specific heatReview of Scientific Instruments, 1983
- A silicon on sapphire thermometer for small sample low temperature calorimetryPhysica B+C, 1981
- Transport properties of polycrystalline silicon filmsJournal of Applied Physics, 1978
- Conduction properties of lightly doped, polycrystalline siliconSolid-State Electronics, 1978
- Specific Heat of Granular Aluminum FilmsPhysical Review B, 1972
- Heat Capacity Measurements on Small Samples at Low TemperaturesReview of Scientific Instruments, 1972
- Impurity Conduction in SiliconPhysical Review B, 1961