Real-Time Digital K-Edge Subtraction Fluoroscopy
- 1 July 1979
- journal article
- Published by Wolters Kluwer Health in Investigative Radiology
- Vol. 14 (4) , 270-278
- https://doi.org/10.1097/00004424-197907000-00002
Abstract
We report in vitro and in vivo trials of K-edge fluoroscopy, by which iodine contrast concentration is displayed live, with tissue and bone images suppressed, free of patient-motion artifacts. Iodine and cerium, 125 and 225 mg/cm2 respectively, filter alternate TV fields of cine-pulsed 50 KVP x-rays. Weighted subtraction of successive TV fields isolates the iodine image and simultaneously minimizes artifacts. Digital techniques are used in real time. At our present x-ray tube limit, 500 mA instantaneous current, the patient exposure is 180 mR/sec and quantum mottle limits the image quality. Integrating four successive difference images provides a compromise between mottle and smoothly moving displays. Cardiovascular images of a 17-kg dog, using 1 ml/kg Renografin-60 injected into a foreleg vein, show that a 15-cm chest thickness is our present practical maximum. This method may be useful in diagnosing cardiovascular anomalies in infants without catheterization or suspension of breathing.Keywords
This publication has 0 references indexed in Scilit: