Effect of Velocity Gradient on the Clotting Time of Blood and on the Consistency of Clots Formed in Vitro
- 1 April 1966
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation Research
- Vol. 18 (4) , 349-356
- https://doi.org/10.1161/01.res.18.4.349
Abstract
Clotting times of human blood and rheology of blood clots have been determined by means of a cone-in-cone viscometer, a ring-in-ring adapter, and a variable-frequency thromboviscometer. The last represents a variable frequency version of the thrombelastograph. The results of clotting time tests on blood samples from 118 donors indicate clearly that clotting time is a function of the velocity gradient at which clotting takes place. Clotting time decreases as the velocity gradient increases. Consistency of blood coagula depends on the velocity gradient at which coagula are formed. Consistency, or viscosity, of coagula decreases when the velocity gradient increases. While the viscosity of clots formed at low rates of shear is of the order of 100 poises, coagula formed at intermediate rates of shear exhibit viscosity of a few decipoises. Viscosity of blood is not significantly altered if coagulation takes place at rates of shear near 400 sec-1.This publication has 4 references indexed in Scilit:
- The influence of the velocity gradient on in vitro blood coagulation and artificial thrombosisJournal of Atherosclerosis Research, 1965
- A TROLLEY VISCOMETER FOR ESTIMATING VISCOSITY AND CLOTTING OF BLOOD IN HOSPITAL WARDSThe Lancet, 1964
- THROMBUS FORMATION IN VITRO: A RHEOLOGICAL AND MORPHOLOGICAL STUDYImmunology & Cell Biology, 1964
- An apparatus for measuring the tensile strength of blood clotsJournal of Clinical Pathology, 1961