Automatic platelet counting with the Coulter particle counter
Open Access
- 1 September 1967
- journal article
- research article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 20 (5) , 777-779
- https://doi.org/10.1136/jcp.20.5.777
Abstract
A method for accurately counting platelets is described using the Coulter counter model B fitted with a standard 100μ aperture tube. This enables the counter to be used for red and white cell as well as platelet counts using the same aperture tube. The method uses standard equipment except for a small inexpensive electronic speed controller.This publication has 5 references indexed in Scilit:
- Platelet Counts with the Coulter CounterAmerican Journal of Clinical Pathology, 1965
- Rapid whole-blood platelet counting using an electronic particle counterJournal of Clinical Pathology, 1963
- Platelet counting using the Coulter electronic counterJournal of Clinical Pathology, 1963
- An automatic platelet counting technique.1962