A technique for the classification and quantification of damage in retrieved total knee prostheses is presented and applied to the examination of 48 removed total condylar‐type knee replacements. The technique involves inspection of all metallic and polyethylene components for evidence of gross deformation, fracture, and damage to articulating surfaces. A grading system was developed to quantitate surface damage on polyethylene components. Results of the examinations are combined with patient variables (weight, activity level, radiographic findings, time of implantation, and results of histology performed on surrounding tissue) to determine correlations between clinical variables and the mechanical damage experienced by the prostheses. For the 48 total condylar‐type prostheses, significant positive correlations were found for the surface damage correlated with the patient's weight and the time the prosthesis was implanted.This publication has 10 references indexed in Scilit:
- Performance of the tibial component in total knee replacement.Journal of Bone and Joint Surgery, 1982
- The effect of carbon fiber reinforcement on contact area, contact pressure, and time‐dependent deformation in polyethylene tibial componentsJournal of Biomedical Materials Research, 1981
- On the true wear rate of ultra high-molecular-weight polyethylene in the total hip prosthesis.Journal of Bone and Joint Surgery, 1980
- Side-effects of prosthetic materials on the human bodyInternational Orthopaedics, 1980
- Wear of high-density polyethylene on bone and cartilageThe Journal of Bone and Joint Surgery. British volume, 1979
- The total condylar knee prosthesis. A report of two hundred and twenty cases.Journal of Bone and Joint Surgery, 1979
- Evaluation of the Tissue Response to the Wear Products of the Hip Joint Endo-ArthroprosthesisBiomaterials, Medical Devices, and Artificial Organs, 1979
- The characteristics of acetabular cups worn in the human bodyThe Journal of Bone and Joint Surgery. British volume, 1978
- The appearances of wear on polyethylene—a comparison of in vivo and in vitro wear surfacesJournal of Biomedical Materials Research, 1978
- A comparison of four models of total knee-replacement prosthesesJournal of Bone and Joint Surgery, 1976