The Pathology of Hancock Standard Porcine Valve Prosthesis: A 20-Year Span of Experience
- 1 December 1990
- journal article
- review article
- Published by Hindawi Limited in Journal of Cardiac Surgery
- Vol. 5 (4) , 328-335
- https://doi.org/10.1111/j.1540-8191.1990.tb00763.x
Abstract
A spectrum of events leading to tissue failure is responsible for late dysfunction of Hancock porcine valve xenografts: (a) Primary failure: dystrophic calcification, thrombosis, fibrous tissue overgrowth, primary tears, cuspal hematomas, and stent postbending. (b) Secondary failure: endocarditis and paravalvular leak. Dystrophic calcification is the main factor influencing long‐term durability and accounts in our experience for 88% of primary failure, through different clinical presentations; particularly, incompetence by cusp tearing and egg‐shell fragmentation is by far the most frequent mode of failure. Cusp degeneration by primary tears (in the absence of dystrophic calcification) is an uncommon event, due to lipid infiltration or to right coronary muscle shelf spontaneous or immuno‐related disruption.Keywords
This publication has 31 references indexed in Scilit:
- Emergency reoperation for primary tissue failure of porcine bioprosthesesThe American Journal of Cardiology, 1987
- Cardiac Valve Prostheses: Pathological and Bioengineering ConsiderationsJournal of Cardiac Surgery, 1987
- Anatomic analysis of removed prosthetic heart valves: Causes of failure of 33 mechanical valves and 58 bioprostheses, 1980 to 1983Human Pathology, 1985
- Calcific degeneration as the main cause of porcine bioprosthetic valve failureThe American Journal of Cardiology, 1984
- Glutaraldehyde-Preserved Porcine BioprosthesisChest, 1983
- Long-term failure rate and morphologic correlations in porcine bioprosthetic heart valvesThe American Journal of Cardiology, 1983
- Calcification of porcine prosthetic heart valves: a radiographic and light microscopy study.Circulation, 1982
- Structure and classification of cuspal tears and perforations in porcine bioprosthetic cardiac valves implanted in patientsThe American Journal of Cardiology, 1981
- Fibrous Tissue Overgrowth on Hancock Mitral Xenografts: A Cause of Late Prosthetic StenosisThe Thoracic and Cardiovascular Surgeon, 1979
- Structural changes in porcine xenografts used as substitute cardiac valves: Gross and histologic observations in 51 glutaraldehyde-preserved Hancock valves in 41 patientsThe American Journal of Cardiology, 1977