Pseudobacteremia Traced to Cross-Contamination by an Automated Blood Culture Analyzer
- 1 February 1984
- journal article
- research article
- Published by Cambridge University Press (CUP) in Infection Control
- Vol. 5 (2) , 75-78
- https://doi.org/10.1017/s0195941700058987
Abstract
Twenty-three episodes of pseudobacteremia occurring over a seven-month period were traced to cross-contamination by the automated blood culture analyzer (BACTEC 460) used in the microbiology laboratory. An epidemiologic investigation of an unusual cluster of three patients with pseudobacteremia caused by oxacillin-resistant Staphylococcus aureus led to the identification of the problem. Mock trials of the blood culture procedure confirmed that the blood culture analyzer was the source of contamination. After the needle sterilizer was replaced by the manufacturer, the problem of cross-contamination abated. Contamination of sterile blood cultures by an instrument intended to identify bacteremia rapidly may lead to incorrect diagnosis, unnecessary administration of antibiotics and prolonged hospitalization. Because of the widespread use of automated blood culture analyzers in the US, physicians, microbiologists, and infection control personnel should be alert to the possibility of cross-contamination and the subtle way in which it may present.This publication has 15 references indexed in Scilit:
- Klebsiella pneumoniaePseudobacteremia Due to Cross-Contamination of A Radiometric Blood Culture AnalyzerInfection Control, 1981
- Pseudobacteremia Attributed to Contamination of Povidone-Iodine with Pseudomonas cepaciaAnnals of Internal Medicine, 1981
- A large outbreak of infections caused by a strain of staphylococcus aureus resistant to oxacillin and aminoglycosidesThe American Journal of Medicine, 1981
- Through A Glass DarklyArchives of internal medicine (1960), 1980
- Nosocomial PseudobacteremiaPublished by American Medical Association (AMA) ,1976
- False-Positive Blood CulturesJAMA, 1976
- Flavobacterium meningosepticum in the hospital environment.Journal of Clinical Pathology, 1976
- Bacterial contamination of blood-drawing equipmentThe Lancet Healthy Longevity, 1973
- Mima polymorpha BacteremiaJAMA, 1972
- PseudosepticemiaAnnals of Internal Medicine, 1969