Multilayer piezoelectric ceramics for two-dimensional array transducers
- 1 September 1994
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 41 (5) , 761-771
- https://doi.org/10.1109/58.308512
Abstract
In medical ultrasound imaging, 2-D array transducers have become essential to implement dynamic focusing and phase-correction in the elevation dimension as well as real-time volumetric scanning. Unfortunately, the small size of a 2-D array element results in a small clamped capacitance and a large electrical impedance near resonance. These elements have poor sensitivity because their impedance is much higher than the electrical impedance of the transmit and receive circuitry. Sensitivity can be improved by using an N layer structure of PZT ceramic with the layers connected acoustically in series and electrically in parallel. For the multilayer ceramic (MLC), the damped capacitance is multiplied by a factor of N/sup 2/ and the electrical impedance by 1/N/sup 2/ compared to a single layer element of the same dimensions. A 3/spl times/43 phased-array transducer has been fabricated using 3 layer PZT-5H material. Each element had a thickness of 0.66 mm and an area of 0.37/spl times/3.5 mm. The MLC was manufactured using thick film technology with plated-through vias to electrically interconnect the electrode layers. The completed transducer was compared to a single layer control array of similar dimensions. With a light epoxy backing and a /spl lambda//4 matching layer, the MLC array elements had an impedance of 100 /spl Omega/ at series resonance of 2.25 MHz, compared to 800 /spl Omega/ for the control elements. The lower impedance of the MLC elements resulted in a minimum round-trip insertion loss of 24.0 dB, compared to an 34.1 dB for the control array elements. These results were consistent with KLM modeling. B-scan images were made of cysts in a tissue-mimicking phantom and of the left kidney in vivo. The images clearly showed a higher signal-to-noise ratio for the MLC array compared to the control. As a result, 2-D arrays made of multilayer ceramics can be used to form images at a higher frequency and greater range than single layer arrays.<>Keywords
This publication has 26 references indexed in Scilit:
- Relaxor ferroelectric materialsPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2002
- Two-dimensional array transducers using thick film connection technologyIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1993
- Multilayer PVDF Polymer Transducers for Pulse Compression ApplicationsPublished by Elsevier ,1993
- Nonlinearity of a composite relaxor ferroelectric transducerPublished by SPIE-Intl Soc Optical Eng ,1992
- In vivo imaging using a copolymer phased arrayUltrasonic Imaging, 1992
- Simulation of piezoelectric devices by two- and three-dimensional finite elementsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1990
- Ultrasound Transducers for Pulse-Echo Medical ImagingIEEE Transactions on Biomedical Engineering, 1983
- On the Generation of High-Frequency Acoustic Energy with Polyvinylidene FluorideIEEE Transactions on Sonics and Ultrasonics, 1980
- Piezoelectric properties of internally electroded PZT multilayersFerroelectrics, 1980
- Equivalent Circuits for Transducers Having Arbitrary Even- Or Odd-Symmetry Piezoelectric ExcitationIEEE Transactions on Sonics and Ultrasonics, 1971