Commissioning and quality assurance of an optically guided three-dimensional ultrasound target localization system for radiotherapy
- 19 July 2002
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 29 (8) , 1781-1788
- https://doi.org/10.1118/1.1494835
Abstract
Recently, there has been proliferation of image-guided positioning systems for high-precision radiation therapy, with little attention given to quality assurance procedures for such systems. To ensure accurate treatment delivery, errors in the imaging, localization, and treatment delivery processes must be systematically analyzed. This paper details acceptance tests for an optically guided three-dimensional (3D) ultrasound system used for patient localization. While all tests were performed using the same commercial system, the general philosophy and procedures are applicable to all systems utilizing image guidance. Determination of absolute localization accuracy requires a consistent stereotactic, or three-dimensional, coordinate system in the treatment planning system and the treatment vault. We established such a coordinate system using optical guidance. The accuracy of this system for localization of spherical targets imbedded in a phantom at depths ranging from 3 to 13 cm was determined to be (average +/- standard deviation) AP = 0.2 +/- 0.7 mm, Lat = 0.9 +/- 0.6 mm, Ax = 0.6 +/- 1.0 mm. In order to test the ability of the optically guided 3D ultrasound localization system to determine the magnitude of an internal organ shift with respect to the treatment isocenter, a phantom that closely mimics the typical human male pelvic anatomy was used. A CT scan of the phantom was acquired, and the regions of interest were contoured. With the phantom on the treatment couch, optical guidance was used to determine the positions of each organ to within imaging uncertainty, and to align the phantom so the plan and treatment machine coordinates coincided. To simulate a clinical misalignment of the treatment target, the phantom was then shifted by different precise offsets, and an experimenter blind to the offsets used ultrasound guidance to determine the magnitude of the shifts. On average, the magnitude of the shifts could be determined to within 1.0 mm along each axis.Keywords
Funding Information
- Whitaker Foundation Biomedical Engineering Research (RG-98-0278)
This publication has 26 references indexed in Scilit:
- Breathing‐synchronized radiotherapy program at the University of California Davis Cancer CenterMedical Physics, 2000
- Accuracy of a photogrammetry‐based patient positioning and monitoring system for radiation therapyMedical Physics, 1999
- 114 Automated target positioning for extracranial radiosurgeryInternational Journal of Radiation Oncology*Biology*Physics, 1999
- Optical High-Precision Three-Dimensional Position Measurement System Suitable for Head Motion Tracking in Frameless Stereotactic RadiosurgeryComputer Aided Surgery, 1998
- Patient setup optimization for external conformal radiotherapyJournal of Image Guided Surgery, 1995
- Photogrammetric accuracy measurements of head holder systems used for fractionated radiotherapyInternational Journal of Radiation Oncology*Biology*Physics, 1994
- An accurate and ergonomic method of registration for image-guided neurosurgeryComputerized Medical Imaging and Graphics, 1994
- The NeurostationTM—A highly accurate, minimally invasive solution to frameless stereotactic neurosurgeryComputerized Medical Imaging and Graphics, 1994
- Coordinate transformations and calculation of the angular and depth parameters for a stereotactic systemMedical Physics, 1987
- Computer calculations of target parameters for a stereotactic apparatusActa Neurochirurgica, 1973