Endovascular Stent-Graft Repair of a Complicated Penetrating Ulcer of the Descending Thoracic Aorta: A Word of Caution
- 1 October 2003
- journal article
- case report
- Published by SAGE Publications in Journal of Endovascular Therapy
- Vol. 10 (5) , 928-931
- https://doi.org/10.1177/152660280301000513
Abstract
Purpose: To report a pitfall encountered during stenting of a complicated penetrating ulcer of the descending thoracic aorta. Case Report: A 65-year-old man was diagnosed with a complicated penetrating ulcer of the thoracic aorta. A 38-mm Talent endograft was implanted. On balloon dilation of the distal end of the endoprosthesis, the terminal bare stent became distorted and penetrated the aortic wall. A 42-mm endoprosthesis was immediately placed to exclude the aortic perforation. Control aortography demonstrated exclusion of the original proximal aortic ulcer and the distal iatrogenic aortic tear. Conclusions: Endoprostheses may present some drawbacks in terms of elasticity and adaptability to tortuous and angulated diseased aortas. Caution is advised in the treatment of penetrating aortic ulcers where the aortic wall is diffusely friable. In this condition, balloon dilation should be limited to the covered portion of the stent-graft to prevent stent distortion and erosion through the aortic wall.Keywords
This publication has 13 references indexed in Scilit:
- Contemporary Management of Acute Aortic DissectionJournal of Endovascular Therapy, 2003
- False Aneurysms at Both Ends of a Descending Thoracic Aortic Stent-Graft: Complication after Endovascular Repair of a Penetrating Atherosclerotic UlcerJournal of Endovascular Therapy, 2003
- Stent graft repair in the aortic arch and descending thoracic aorta: A 4-year experienceJournal of Vascular Surgery, 2002
- Endovascular stent-graft repair of complicated penetrating atherosclerotic ulcers of the descending thoracic aortaJournal of Vascular Surgery, 2002
- Endovascular Repair for Acute Rupture of the Descending Thoracic AortaJournal of Endovascular Therapy, 2002
- Endovascular Stent-Graft Treatment of a Ruptured Thoracic Aortic UlcerJournal of Endovascular Therapy, 2002
- Stent-Graft Treatment of Penetrating Thoracic Aortic UlcersJournal of Endovascular Therapy, 2002
- Endovascular Treatment of Aortic Type B Dissection and Penetrating Ulcer Using Commercially Available Stent-GraftsAmerican Journal of Roentgenology, 2001
- Endovascular Repair of a Non-Contained Aortic Rupture Caused by a Penetrating Aortic UlcerCardioVascular and Interventional Radiology, 2001
- The “first generation” of endovascular stent-grafts for patients with aneurysms of the descending thoracic aortaThe Journal of Thoracic and Cardiovascular Surgery, 1998