Response Function of NaI(Tl) Scintillation Counters
- 1 May 1956
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 27 (5) , 269-270
- https://doi.org/10.1063/1.1715539
Abstract
The response function of NaI(Tl) scintillation detectors for gamma radiation has been calculated by the Monte Carlo method. Summary results of this calculation are presented in terms of the photofraction, i.e., the ratio of the area under the photopeak to the total area of a scintillation spectrum. Values of the photofraction are given for collimated monoenergetic radiation with energies between 0.279 and 4.45 Mev incident centrally on the endface of cylindrical crystals ranging in size from 0.25 in. (radius)×0.5 in. (length) to 2.5×9 in.Keywords
This publication has 4 references indexed in Scilit:
- Capture to Positron Ratio forPhysical Review B, 1954
- Scintillation Spectrometers for Measuring the Total Energy of X-Ray PhotonsReview of Scientific Instruments, 1954
- The Albedo of Various Materials for 1-Mev PhotonsPhysical Review B, 1954
- Energy Resolution of ?-Radiation up to 18 MeV by Sodium Iodide Scintillation CountersAustralian Journal of Physics, 1953