An Amperometric Needle‐type Glucose Sensor Tested in Rats and Man
- 1 April 1988
- journal article
- Published by Wiley in Diabetic Medicine
- Vol. 5 (3) , 248-252
- https://doi.org/10.1111/j.1464-5491.1988.tb00978.x
Abstract
An amperometric glucose-measuring 25 gauge (0.5 mm diameter) needle-type sensor has been developed using a glucose oxidase and dimethyl ferrocene paste behind a semi-permeable membrane situated over a window in the needle. Electron transfer results in direct current generation. Sensors have been tested subcutaneously in the abdomen both in anaesthetized rats (40 sensors, 11 rats) and in normal, conscious man (20 sensors, 10 subjects). In rats the blood glucose was modulated by glucose and by insulin infusion. In man the glucose concentrations were rapidly changed by use of a glucose clamp at 12 mmol/l plasma concentration for 2 h, after which the glucose returned to normal. In rats the median correlation between glucose change was 0.83 with an interquartile range from 0.70 to 0.92, and in man the median correlation was 0.80 with an interquartile range 0.67 to 0.86. Hysteresis, a measure of the accuracy on the upswing and downswing, was not a problem and cross-correlation showed no phase-lag. There were quantitative differences between in vitro calibration and the performance in vivo, reflecting the different conditions of use. The current in response to a glucose concentration was stable over 6.0 h in rats and 4.5 h in manKeywords
This publication has 16 references indexed in Scilit:
- Subcutaneous implantation of a ferrocene-mediated glucose sensor in pigsDiabetologia, 1986
- Ferrocene-mediated enzyme electrode for amperometric determination of glucoseAnalytical Chemistry, 1984
- Comparative toxicity of the hematinic MDL 80,478 — Effects on the liver and adrenal cortex of the dog, rat and monkeyFundamental and Applied Toxicology, 1983
- WEARABLE ARTIFICIAL ENDOCRINE PANCREAS WITH NEEDLE-TYPE GLUCOSE SENSORThe Lancet, 1982
- Implanted Electroenzymatic Glucose SensorsDiabetes Care, 1982
- Potentiometric Measurement of Glucose Concentration with an Immobilized Glucose Oxidase/Catalase ElectrodeDiabetes Care, 1982
- Continuous subcutaneous insulin infusion: an approach to achieving normoglycaemia.BMJ, 1978
- The Analysis of Time Series: Theory and PracticePublished by Springer Nature ,1975
- Chronic toxicity of dicyclopentadienyliron (ferrocene) in dogsToxicology and Applied Pharmacology, 1969
- ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERYAnnals of the New York Academy of Sciences, 1962