Size of the detector in confocal imaging systems
- 1 April 1987
- journal article
- Published by Optica Publishing Group in Optics Letters
- Vol. 12 (4) , 227-229
- https://doi.org/10.1364/ol.12.000227
Abstract
We consider the role of the detector size in the imaging of confocal systems. We introduce a criterion for detector size to give true confocal operation and show that the lateral resolution is considerably more sensitive to detector size than is the depth-discrimination or optical-sectioning property.Keywords
This publication has 4 references indexed in Scilit:
- Experimental observations of the depth-discrimination properties of scanning microscopesOptics Letters, 1981
- Depth of field in the scanning microscopeOptics Letters, 1978