Blood Flow Pulsatility Effects upon Oxygen Transfer in Artificial Lungs
- 1 November 2003
- journal article
- research article
- Published by Wolters Kluwer Health in Asaio Journal
- Vol. 49 (6) , 678-686
- https://doi.org/10.1097/01.mat.0000094041.69991.0d
Abstract
This report discusses theoretical effects of blood flow pulsatility upon the rate of oxygen transfer in artificial lungs, demonstrates the effects with in vitro tests upon commercial oxygenators, and applies the theory to these oxygenators and to a thoracic artificial lung. Steady flow gas transfer theory is applied to pulsatile flow by using the instantaneous value of flow rate at each instant of time, that is, quasi-steady gas transfer. The theory suggests that the local rate of oxygen transfer for a given device and blood composition is proportional to the flow rate to a power less than unity and to the hemoglobin saturation level. It predicts, for some cases, overall reduced rates of gas transfer for pulsatile flow relative to those at steady flow for the same mean blood flow rates. In vitro bovine blood tests, using pediatric oxygenators, a pulsatile pump, and an adjustable compliance chamber, indicate a significant average 10% reduction of oxygen transfer for pulsatile flow relative to steady flow. The application of the theory to the oxygenators predicts gas transfer values that are in agreement with those measured during the experiments. The results have implications in the design of implantable thoracic artificial lungs, which should include a compliant section to dampen the cardiac pulse. A relatively small compliance (0.2 ml/mm Hg) at the thoracic artificial lung inlet is sufficient to obtain approximately 95% of steady flow oxygen transfer.Keywords
This publication has 7 references indexed in Scilit:
- Hemodynamic Effects of Attachment Modes and Device Design of a Thoracic Artificial LungAsaio Journal, 2000
- TOTAL RESPIRATORY SUPPORT WITH A COMPLIANT ARTIFICIAL LUNG PERFUSED BY THE RIGHT HEARTAsaio Journal, 1999
- Mass transfer efficiency of a commercial hollow fibre oxygenator during six-hour in vitro perfusion with steady and with pulsatile blood flowThe International Journal of Artificial Organs, 1998
- Testing of an Intrathoracic Artificial Lung in a Pig ModelPublished by Wolters Kluwer Health ,1996
- Use of a Mathematical Model to Predict Oxygen Transfer Rates in Hollow Fiber Membrane OxygenatorsAsaio Journal, 1994
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Landé-Edwards membrane oxygenator without pump as total lung prosthesisThe Journal of Thoracic and Cardiovascular Surgery, 1975