Influence of droplet size, capillary–cone distance and selected instrumental parameters for the analysis of noncovalent protein–ligand complexes by nano‐electrospray ionization mass spectrometry
- 7 September 2004
- journal article
- research article
- Published by Wiley in Journal of Mass Spectrometry
- Vol. 39 (9) , 1059-1067
- https://doi.org/10.1002/jms.685
Abstract
It has been suggested in the literature that nano-electrospray ionization (nano-ESI) mass spectrometry better reflects the equilibrium between complex and free protein in solution than pneumatically assisted electrospray ionization (ESI) in noncovalent interaction studies. However, no systematic studies of the effects of ionization conditions have been performed to support this statement. In the present work, different instrumental and sample-derived parameters affecting the stability of noncovalent complexes during analysis by nano-ESI were investigated. In general, increased values of parameters such as drying gas flow-rate, ion-source temperature, capillary tip voltage and buffer concentration lead to a dissociation of ribonuclease A (RNAse)–cytidine 2′-monophosphate (CMP) and cytidine 5′-triphosphate (CTP) complexes. The size of the electrosprayed droplets was shown to be an important issue. Increasing the capillary to cone distance yielded an increased complex to free protein ratio when a hydrophilic ligand was present and the reverse effect was obtained with a hydrophobic ligand. Important in this regard is the degree of sampling of ions originating from late-generation residue droplets, that is, ions present in the droplet bulk. Sampling of these ions increases with longer capillary–cone distance (flight time). Furthermore, when the sample flow-rate was increased by increasing the capillary internal tip i.d. from 4 to 30 µm, a decreased complex to free protein ratio for the RNAse–CTP system was observed. This behavior was consistent with the change in surface to volume ratio for flow-rates between 2 and 100 nl min−1. Finally, polarity switching between positive and negative ion modes gave a higher complex to free protein ratio when the ligand and the protein had the same polarity. Copyright © 2004 John Wiley & Sons, Ltd.Keywords
This publication has 34 references indexed in Scilit:
- Use of electrospray ionization mass spectrometry to study binding interactions between a replication terminator protein and DNAProtein Science, 2009
- Automated Nano-Electrospray Mass Spectrometry for Protein-Ligand Screening by Noncovalent Interaction Applied to Human H-FABP and A-FABPSLAS Discovery, 2003
- Protein complexes gain momentumCurrent Opinion in Structural Biology, 2002
- On‐line microdialysis for enhanced resolution and sensitivity during electrospray mass spectrometry of non‐covalent complexes and competitive binding studiesRapid Communications in Mass Spectrometry, 2002
- Advantages and drawbacks of nanospray for studying noncovalent protein–DNA complexes by mass spectrometryRapid Communications in Mass Spectrometry, 2002
- Probing metal ion binding and conformational properties of the colicin E9 endonuclease by electrospray ionization time‐of‐flight mass spectrometryProtein Science, 2002
- Electrospray Ionization Mass Spectrometry: A Promising New Technique in the Study of Protein/DNA Noncovalent ComplexesBiochemical and Biophysical Research Communications, 1999
- Probing the Nature of Noncovalent Interactions by Mass Spectrometry. A Study of Protein−CoA Ligand Binding and AssemblyJournal of the American Chemical Society, 1996
- Electrospray and Taylor-Cone theory, Dole's beam of macromolecules at last?International Journal of Mass Spectrometry and Ion Processes, 1994
- Detection of noncovalent receptor-ligand complexes by mass spectrometryJournal of the American Chemical Society, 1991