A convenient on‐line device for reagent addition, sample mixing, and temperature control of cell suspensisons in flow cytometry
- 1 January 1985
- Vol. 6 (1) , 69-73
- https://doi.org/10.1002/cyto.990060113
Abstract
We describe a simple and inexpensive device which permits the addition of up to three different solutions into a cell suspension which is on‐line in a flow cytometer. The mixing chamber houses a disposable plastic cuvette stirred with a magnetic stirrer. The sample chamber is attached to a circulating water bath, hence accurate temperature control is achieved. Because the system is prepressurized and the sample line is very short, the delay time‐between the point of sample modification and the point of analysis is reduced to a few seconds. Thus reagents may be added rapidly, and kinetic measurements of high temporal resolution are possible. Because the temperature of the sample chamber is regulated, binding can be observed over longer time periods than was previously possible. We demonstrate the usefulness of this device in determining the binding of fluoresceinated hexapeptide to human neutrophils under conditions where the stimulus is infused into the cell suspension while on‐line in the cytometer.Keywords
This publication has 8 references indexed in Scilit:
- Neutrophil degranulation detected by right angle light scattering: spectroscopic methods suitable for simultaneous analyses of degranulation or shape change, elastase release, and cell aggregation.The Journal of Immunology, 1984
- The dynamics of ligand-receptor interactions. Real-time analyses of association, dissociation, and internalization of an N-formyl peptide and its receptors on the human neutrophil.Journal of Biological Chemistry, 1984
- Is Cytosolic Ionized Calcium Regulating Neutrophil Activation?Science, 1983
- Ligand/receptor internalization: A kinetic, flow cytometric analysis of the internalization of N‐formyl peptides by human neutrophilsCytometry, 1983
- Flow cytometric studies of oxidative product formation by neutrophils: a graded response to membrane stimulation.The Journal of Immunology, 1983
- Analysis of ligand‐receptor interactions with the fluorescence activated cell sorterCytometry, 1982
- A new flow cytometric transducer for fast sample throughput and time resolved kinetic studies of biological cells and other particlesCytometry, 1982
- Flow cytofluorometric analysis of insulin binding and internalization by swiss 3T3 cellsCytometry, 1982